image

8.4. Analytical performance of N Latex FLC and Freelite assays

Chapter 8

The FLC assays most frequently used as an alternative to Freelite are the Siemens N Latex FLC assays. These assays have been used in a number of evaluation and comparison studies, and are described below.

8.4.1. Calibration

An international reference material is not available for FLCs, therefore, master standards for the N Latex FLC assays were calibrated against the Freelite assays [199]. However, the absolute values returned by the two assays do not compare well. This is discussed further in Section 8.5.3.

8.4.2. Precision

As the measuring ranges of Freelite and N Latex FLC assays are different, samples used in precision studies must be suitably selected to ensure a fair comparison. This was illustrated in a precision study by Pretorius et al. [175] in which the samples with the lowest concentrations were well within the standard N Latex FLC measuring range (above the second calibrator point), but at or below the standard Freelite measuring range. Unsurprisingly, the total %CV of the lowest sample for the Freelite assays was >20%, but <7% for N Latex FLC (Figure 8.3.)

A better-designed precision study by Lock et al. [201] used samples that either fell within the reference range or were moderately elevated by both FLC assays. The authors concluded that the within-batch and between-batch precision of both Freelite and N Latex FLC assays was acceptable (Table 8.3). A report by Sharrod-Cole et al. [202] reached similar conclusions. Finally, Carr-Smith et al. [921] used UK NEQAS external quality control data to compare the inter-laboratory variation of Freelite assays on the SPAPLUS® with that of N Latex FLC assays on the Siemens BN™II. The authors conclude that the performance of Freelite was at least equivalent, and often superior, to N Latex FLC for the majority of distributions (Section 39.3).

Maintaining batch-to-batch consistency is an essential requirement for any FLC assay, which may be used to monitor patients over the course of many years. Freelite polyclonal antisera are produced using standardised manufacturing procedures that include careful blending to achieve balanced, reproducible recognition of FLC molecules, with good between-batch consistency. This is further discussed in Section 5.5.

Assay Precision fluid Freelite (%CV) N Latex FLC (%CV)
Within-batch precisionκ sFLCNormal 9.8 2.2
λ sFLCNormal 7.6 1.8
Between-batch precisionκ sFLCNormal 7.1 5.1
Elevated 5.8 6.4
λ sFLCNormal 7.0 3.6
Elevated 6.4 5.6

Table 8.3. Precision of Freelite and N Latex FLC assays [201]. Reproducibility data based on 10 replicates, except N Latex FLC between-batch which was based on 3 replicates.

A theoretical advantage of monoclonal over polyclonal antibodies is reduced variation between reagent lots. Pretorius et al. [175] assessed variance between two N Latex FLC reagent lots, calibrated with the same material. Whilst the κ N Latex FLC assay demonstrated negligible bias (-1.7%), the λ N Latex FLC assay demonstrated significant bias (-5.4%). For both κ and λ N Latex FLC assays the difference in values for a number of samples was >20% (Figure 8.4). This included one notable extreme value which showed a >80% difference in λ sFLC concentration between reagent lots (Figure 8.4B). This indicates that use of monoclonal antibodies does not eliminate between-batch variation.

8.4.3. Linearity

Occasional sample non-linearity (i.e. when a sample measured at different dilutions gives a substantially different result) occurs with most immunoassays. Examples of sample non-linearity with Freelite assays are described and discussed in Section 7.3. Pretorius et al. [175] reported that sample non-linearity occurred in approximately half of all monoclonal and polyclonal samples, with both Freelite and N Latex FLC assays. The authors concluded that non-linearity of FLC assays is a property of the individual sample and not a method-specific phenomenon. Freelite assays undergo validation during development (based on protocols defined by the Clinical and Laboratory Standards Institute) to ensure that the assays are linear (Section 5.6).

8.4.4. Antigen excess

Antigen excess causes immunoassays to underestimate high concentrations of protein (Section 7.4). Freelite antigen excess checks can be performed on all platforms, and are carried out in one of two ways: 1) automatic prozone parameters, e.g. Binding Site SPAPLUS® and Optilite®; or 2) a sample dilution protocol detailed in the product insert, e.g. Siemens BN™II. Freelite antigen excess is further discussed in Section 7.4.

The Siemens N Latex FLC assays include antigen excess detection parameters based on a pre-reaction protocol. For the pre-reaction, a small volume of sample is initially added, followed by the rest of the sample for the primary reaction. If the signal at the end of the pre-reaction is above a certain threshold value, the instrument automatically performs a higher dilution. te Velthuis et al. [199] described the initial validation of N Latex FLC antigen excess parameters. In that study, over 2,000 serum samples were screened to identify the two with the highest concentration of κ or λ sFLCs: 23,000 and 57,000 mg/L, respectively. Serial dilutions of these two high samples failed to demonstrate antigen excess. However, this protocol does not provide a good assessment of antigen excess susceptibility because 1) only two samples were tested; and 2) they were chosen in a manner that was most likely to select samples not subject to antigen excess.

Jacobs et al. [203] compared the results from Freelite and N Latex FLC assays for 46 samples containing polyclonal or monoclonal sFLCs and using multiple sample dilutions. For monoclonal κ FLCs measured by Freelite or N Latex FLC, the discrepancies of concentrations between different dilutions were sometimes large. In addition, λ sFLC concentrations were underestimated at higher concentrations by both methods. However, this study failed to give a clear definition of non-linearity or antigen excess and the incidence of neither phenomenon was reported.

Two further studies compared the incidence of antigen excess and gross non-linearity for Freelite and Siemens N Latex FLC assays by testing samples at higher sample dilutions, on the BNII [204][205]. Harding et al. [205] reported that κ sFLC antigen excess was observed in 3/91 (3%) samples by Freelite and 4/91 (4%) by N Latex FLC; all cases of Freelite antigen excess were correctly identified using the manufacturer’s dilution protocol. λ sFLC antigen excess was not observed with Freelite assays, in keeping with previous reports that suggest that these λ assays are less prone to antigen excess than κ assays (Section 7.4.1). In contrast, 3/91 (3%) of samples showed λ sFLC antigen excess with the N Latex FLC assay. Burden et al. [204] reported that a proportion of samples measured with κ or λ N Latex FLC assays gave results that were 5-9 times greater at the higher sample dilution compared with the result at the initial dilution (Table 8.4); this is indicative of antigen excess despite the fact that the assays should have automatic protection from antigen excess issues.

N Latex FLC Freelite
Dilution κ FLC (mg/L) Dilution κ FLC (mg/L)
1:100 71:100 >171
1:400 NR 1:400 >684
1:2000 2891:2000 617
N Latex FLC Freelite
Dilution λ FLC (mg/L) Dilution λ FLC (mg/L)
1:20 41:100 >155
1:100 61 1:400 >620
1:400 1581:2000 1020

Table 8.4. sFLC concentrations measured by N Latex FLC and Freelite assays at several sample dilutions [204]. NR: not reported.

Therefore, it should be noted that: 1) both Freelite and N Latex FLC assays show antigen excess in a small number of samples; 2) antigen excess parameters utilised in the N-Latex assays do not prevent all samples from exhibiting antigen excess; and 3) discrepant results between the two assays could be due to antigen excess. This third point is discussed further in Section 8.5.6.

Figures

References